FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2201265 · Received August 3, 2011

Report

Report Number
1720753-2011-20661
Event Type
Malfunction
Date Received
August 3, 2011
Date of Event
July 27, 2011
Report Date
August 3, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REP ADJUSTED THE 5VDC POWER SUPPLY. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD ONLY PRODUCE INTERMITTENT FLUOROSCOPIC X-RAY. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1