FDA Adverse Event Malfunction Summary report: N

CARESCAPE MONITOR B850

MDR report key: 2201199 · Received July 29, 2011

Report

Report Number
2124823-2011-00125
Event Type
Malfunction
Date Received
July 29, 2011
Date of Event
July 4, 2011
Report Date
July 29, 2011
Manufacturer
GE HEALTHCARE
Product Code
MHX
PMA / PMN Number
K092027
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PT USE THE CARESCAPE B850 PT MONITOR EMITTED AN UNUSUAL CONSTANT BEEP-BEEP TONE THAT DID NOT AUDIBLY CORRESPOND TO AN EXPECTED ALARM TONE OR OTHER RECOGNIZABLE TONE, AND DID NOT CORRELATE TO ANY VISUAL PT CONDITION. AS LONG AS THIS TONE IS SOUNDING NO OTHER DISCERNIBLE AUDIBLE ALARMS WILL BE ANNOUNCED. VISUAL ALARMS WERE NOT AFFECTED. THERE WAS NO DEATH OR SERIOUS INJURY ASSOCIATED WITH THIS EVENT, NOR AN INDICATION THAT INTERVENTION WAS PROVIDED. GE HEALTHCARE'S INVESTIGATION IS ONGOING. A F/U REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CARESCAPE MONITOR B850 MULTI-PARAMETER PATIENT MONITOR MHX GE HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1