FDA Adverse Event
Malfunction
Summary report: N
CARESCAPE MONITOR B850
MDR report key: 2201195
·
Received July 29, 2011
Report
- Report Number
- 2124823-2011-00119
- Event Type
- Malfunction
- Date Received
- July 29, 2011
- Date of Event
- July 4, 2011
- Report Date
- July 29, 2011
- Manufacturer
- GE HEALTHCARE
- Product Code
- MHX
- PMA / PMN Number
- K092027
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING PT USE THE CARESCAPE B850 PT MONITOR EMITTED AN UNUSUAL CONSTANT BEEP-BEEP TONE THAT DID NOT AUDIBLY CORRESPOND TO AN EXPECTED ALARM TONE OR OTHER RECOGNIZABLE TONE, AND DID NOT CORRELATE TO ANY VISUAL PT CONDITION. AS LONG AS THIS TONE IS SOUNDING NO OTHER DISCERNIBLE AUDIBLE ALARMS WILL BE ANNOUNCED. VISUAL ALARMS WERE NOT AFFECTED. THERE WAS NO DEATH OR SERIOUS INJURY ASSOCIATED WITH THIS EVENT, NOR AN INDICATION THAT INTERVENTION WAS PROVIDED. GE HEALTHCARE'S INVESTIGATION IS ONGOING. A F/U REPORT WILL BE SUBMITTED ONCE THE INVESTIGATION HAS BEEN COMPLETED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CARESCAPE MONITOR B850 | MULTI-PARAMETER PATIENT MONITOR | MHX | GE HEALTHCARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |