FDA Adverse Event Malfunction Summary report: N

HERCULES 3 STAGE BALLOON ESOPHAGEAL

MDR report key: 2201194 · Received July 29, 2011

Report

Report Number
1037905-2011-00496
Event Type
Malfunction
Date Received
July 29, 2011
Date of Event
June 21, 2011
Report Date
June 29, 2011
Manufacturer
WILSON-COOK MEDICAL INC
Product Code
KNQ
PMA / PMN Number
K061937
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

EVAL: THE PRODUCT SAID TO BE INVOLVED WAS RETURNED FOR EVAL AND WE CONFIRMED ALL SECTIONS OF THE DEVICE ARE PRESENT. THE LABEL OF THE OPENED PACKAGING INCLUDED IN THE RETURN MATCHES THE PRODUCT RETURNED. DURING A VISUAL EXAMINATION OF THE BALLOON MATERIAL, WE CONFIRMED THE PRESENCE OF A SMALL SPLIT IN THE BALLOON MATERIAL. THE SPLIT IS POSITIONED LENGTH-WISE ON THE BALLOON AND IS APPROX 1CM. THE DAMAGED AREA IS LOCATED NEAR THE PROXIMAL END. THE CATHETER TUBING DOES NOT EXHIBIT ANY STRETCHED OR DAMAGED AREAS. DURING A FUNCTIONAL EVAL OF THE PRODUCT SAID TO BE INVOLVED, A COOK QUANTUM INFLATION DEVICE FILLED WITH WATER WAS USED IN AN ATTEMPT TO INFLATE THE BALLOON. WHEN THE BALLOON IS INFLATED WITH WATER, WATER EXITS THE BALLOON MATERIAL AT THE LOCATION OF THE LARGE SPLIT. THE BALLOON WILL NOT HOLD PRESSURE AND CANNOT PERFORM DILATION IN THIS CONDITION. A MFG DISCREPANCY OR ANOMALY THAT COULD HAVE CONTRIBUTED TO THIS REPORT WAS NOT OBSERVED DURING OUR LAB EVAL. THE DEVICE HISTORY RECORD FOR THE LOT NUMBER SAID TO BE INVOLVED WAS REVIEWED. A DISCREPANCY OR ANOMALY WAS OBSERVED WITH THE PRODUCT THAT WAS RELEASED FOR DISTRIBUTION. A SAMPLE TEST FROM THIS LOT COULD NOT BE CONDUCTED BECAUSE NONE OF THE PRODUCT FROM THIS LOT REMAINED IN INVENTORY. CONCLUSIONS: A DEFINITIVE CAUSE FOR THIS OBSERVATION COULD NOT BE DETERMINED BECAUSE THE ACTUAL PRODUCT HANDLING CONDITIONS COULD NOT BE DUPLICATED IN THE LAB SETTING. THIS LIMITS OUR ABILITY TO CONCLUSIVELY DETERMINE A CAUSE. A POSSIBLE CONTRIBUTING FACTOR TO BALLOON MATERIAL RUPTURE IS INADEQUATE LUBRICATION OF THE BALLOON WITH A LUBRICATING AGENT. THE INSTRUCTIONS FOR USE DIRECT THE USER TO APPLY A LUBRICATING AGENT TO THE BALLOON TO FACILITATE PASSAGE THROUGH THE ENDOSCOPE ACCESSORY CHANNEL. THIS ACTIVITY WILL AID IN ENDOSCOPIC ADVANCEMENT AND BALLOON PRESERVATION. THE INSTRUCTIONS FOR USE DIRECT THE USER TO APPLY NEGATIVE PRESSURE TO THE CATHETER TO FACILITATE PASSAGE THROUGH THE ENDOSCOPE. THE APPLICATION OF NEGATIVE PRESSURE WILL ASPIRATE ALL RESIDUAL AIR FROM THE BALLOON AND EASE ENDOSCOPIC ADVANCEMENT. NEGATIVE PRESSURE WILL ALSO AID IN BALLOON PRESERVATION AND OPTIMIZE BALLOON PERFORMANCE. A BALLOON RUPTURE CAN OCCUR IF THE BALLOON IS INFLATED PRIOR TO ADVANCEMENT THROUGH THE ENDOSCOPE OR IF THE BALLOON IS INFLATED WHILE PARTIALLY OR FULLY INSIDE THE ACCESSORY CHANNEL OF THE ENDOSCOPE. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING PRECAUTION: "DO NOT PRE-INFLATE THE BALLOON." THE INSTRUCTIONS FOR USE DIRECT THE USER TO ENSURE THE BALLOON IS COMPLETELY VISUALIZED AND POSITIONED BEFORE INFLATION. ANOTHER POSSIBLE CONTRIBUTING FACTOR IS USING A COMPROMISED INFLATION DEVICE TO INFLATE THE BALLOON. IF THE PRESSURE READING OF THE INFLATION DEVICE IS INACCURATE, THIS COULD CONTRIBUTE TO OVER INFLATION, POSSIBLY RESULTING IN A RUPTURE OF THE BALLOON MATERIAL. THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING WARNING: "DURING DILATION, DO NOT INFLATE BALLOON BEYOND THE MAXIMUM INDICATED INFLATION PRESSURE." OVER INFLATION CAN CAUSE DAMAGE TO THE BALLOON DILATOR, SUCH AS A RUPTURE OR SPLIT. PRIOR TO DISTRIBUTION, ALL HERCULES 3 STAGE ESOPHAGEAL BALLOON DILATORS ARE SUBJECTED TO A VISUAL EXAMINATION TO ENSURE DEVICE INTEGRITY. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE LOT SAID TO BE INVOLVED MET ALL MFG REQUIREMENTS PRIOR TO SHIPMENT. CORRECTIVE ACTION: BASED ON THE QUALITY ENGINEERING RISK ASSESSMENT NO CORRECTIVE ACTION IS WARRANTED AT THIS TIME. QUALITY ASSURANCE WILL CONTINUE TO MONITOR FOR COMPLAINT TRENDS.

Description of Event or Problem · 1

DURING AN UPPER ESOPHAGEAL DILATION PROCEDURE, THE PHYSICIAN USED A COOK HERCULES 3 STAGE BALLOON. THE PROXIMAL SECTION OF THE BALLOON BEGAN TO LEAK DURING USE. ANOTHER BALLOON WAS USED TO COMPLETE THE PROCEDURE. A SECTION OF THE DEVICE DID NOT REMAIN INSIDE THE PT'S BODY. THE PT DID NOT REQUIRE ANY ADDITIONAL PROCEDURES DUE TO THIS OCCURRENCE. ACCORDING TO THE INITIAL REPORTER, THE PT DID NOT EXPERIENCE ANY ADVERSE EFFECTS DUE TO THIS OCCURRENCE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HERCULES 3 STAGE BALLOON ESOPHAGEAL KNQ, ESOPHAGEAL DILATOR KNQ WILSON-COOK MEDICAL INC W2938936

Patients

Seq Age Sex Outcome Treatment
1 OLYMPUS GIF ENDOSCOPE| (B)(4) (COOK DILATION SYRINGE)