FDA Adverse Event Malfunction Summary report: N

EVIS EXERA III COLONOVIDEOSCOPE

MDR report key: 22011382 · Received May 13, 2025

Report

Report Number
9610595-2025-08449
Event Type
Malfunction
Date Received
May 13, 2025
Date of Event
May 9, 2025
Report Date
May 13, 2025
Manufacturer
AIZU OLYMPUS CO., LTD.
Product Code
FDF
PMA / PMN Number
K131780
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: THE USE OF PRODUCTS THAT CONTAIN SILICONE CAN CAUSE DEPOSITS OF WHITE FOREIGN MATERIAL. SILICONE IS INCLUDED IN SIMETHICONE, A DEFOAMING AGENT, LUBRICANTS. SIMETHICONE AND PETROLEUM/OIL/SILICONE-BASED LUBRICANTS ARE NON-WATER SOLUBLE AND THUS NOT RECOMMENDED FOR USE BY OLYMPUS. THE EVENT CAN BE PREVENTED BY FOLLOWING THE INSTRUCTIONS FOR USE WHICH STATE: ¿NOTHING OTHER THAN STERILE WATER SHOULD BE USED FOR AIR/WATER FEEDING. NO ADDITIVES SHOULD BE PUT INTO THE STERILE WATER. NON-STERILE WATER MAY CAUSE PATIENT CROSS-CONTAMINATION AND/OR INFECTION.¿ SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS OBSERVED DURING THE DEVICE EVALUATION, THAT THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN AIR/WATER TUBE AND THE JET TUBE. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
967382 EVIS EXERA III COLONOVIDEOSCOPE COLONOVIDEOSCOPE FDF AIZU OLYMPUS CO., LTD. CF-H190L

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown