EVIS EXERA III COLONOVIDEOSCOPE
Report
- Report Number
- 9610595-2025-08449
- Event Type
- Malfunction
- Date Received
- May 13, 2025
- Date of Event
- May 9, 2025
- Report Date
- May 13, 2025
- Manufacturer
- AIZU OLYMPUS CO., LTD.
- Product Code
- FDF
- PMA / PMN Number
- K131780
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- 003
Narratives
THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: THE USE OF PRODUCTS THAT CONTAIN SILICONE CAN CAUSE DEPOSITS OF WHITE FOREIGN MATERIAL. SILICONE IS INCLUDED IN SIMETHICONE, A DEFOAMING AGENT, LUBRICANTS. SIMETHICONE AND PETROLEUM/OIL/SILICONE-BASED LUBRICANTS ARE NON-WATER SOLUBLE AND THUS NOT RECOMMENDED FOR USE BY OLYMPUS. THE EVENT CAN BE PREVENTED BY FOLLOWING THE INSTRUCTIONS FOR USE WHICH STATE: ¿NOTHING OTHER THAN STERILE WATER SHOULD BE USED FOR AIR/WATER FEEDING. NO ADDITIVES SHOULD BE PUT INTO THE STERILE WATER. NON-STERILE WATER MAY CAUSE PATIENT CROSS-CONTAMINATION AND/OR INFECTION.¿ SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.
IT WAS OBSERVED DURING THE DEVICE EVALUATION, THAT THE COLONOVIDEOSCOPE EXHIBITED FOREIGN OBJECTS IN AIR/WATER TUBE AND THE JET TUBE. THERE WAS NO PATIENT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 967382 | EVIS EXERA III COLONOVIDEOSCOPE | COLONOVIDEOSCOPE | FDF | AIZU OLYMPUS CO., LTD. | CF-H190L |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |