Summary report: N
CONTOUR TEST STRIPS
MDR report key: 2201084
·
Received July 1, 2011
Report
- Report Number
- 1826988-2011-00381
- Event Type
- Malfunction
- Date Received
- July 1, 2011
- Date of Event
- June 1, 2011
- Report Date
- June 13, 2011
- Manufacturer
- BAYER HEALTHCARE, LLC
- Product Code
- NBW
- PMA / PMN Number
- K062058
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE ADVOCATE STATED THAT SHE RECEIVED BLOOD GLUCOSE READINGS OF 300 AND 280 MG/DL USING HER CONTOUR METER AND A READING OF 142 MG/DL USING THE CUSTOMER'S CONTOUR METER. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENT WAS ALLEGED. THE ADVOCATE WAS ADVISED TO RETURN THE TEST STRIPS FOR EVAL. REPLACEMENT TEST STRIPS AND A NEW METER WERE SENT TO THE ADVOCATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CONTOUR TEST STRIPS | NBW | BAYER HEALTHCARE, LLC | 7098C | 0J3C07 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |