FDA Adverse Event Malfunction Summary report: N

CONTOUR TEST STRIPS

MDR report key: 2201084 · Received July 1, 2011

Report

Report Number
1826988-2011-00381
Event Type
Malfunction
Date Received
July 1, 2011
Date of Event
June 1, 2011
Report Date
June 13, 2011
Manufacturer
BAYER HEALTHCARE, LLC
Product Code
NBW
PMA / PMN Number
K062058
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE ADVOCATE STATED THAT SHE RECEIVED BLOOD GLUCOSE READINGS OF 300 AND 280 MG/DL USING HER CONTOUR METER AND A READING OF 142 MG/DL USING THE CUSTOMER'S CONTOUR METER. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT. NO ADVERSE EVENT WAS ALLEGED. THE ADVOCATE WAS ADVISED TO RETURN THE TEST STRIPS FOR EVAL. REPLACEMENT TEST STRIPS AND A NEW METER WERE SENT TO THE ADVOCATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONTOUR TEST STRIPS NBW BAYER HEALTHCARE, LLC 7098C 0J3C07

Patients

Seq Age Sex Outcome Treatment
1 UNK