FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 2201082 · Received July 1, 2011

Report

Report Number
3004209178-2011-81994
Event Type
Malfunction
Date Received
July 1, 2011
Date of Event
June 15, 2011
Report Date
June 15, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AR, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

THE CUSTOMER REPORTED A MOTOR ERROR ALARM DURING THE REWIND PROCESS. THE INSULIN PUMP WAS NOT EXPOSED TO A HIGH MAGNETIC FIELD. DURING THE CALL, THE CUSTOMER NOTICED THAT INSULIN LEAKED FROM THE RESERVOIR INTO THE RESERVOIR COMPARTMENT. ADVISED THE CUSTOMER THAT THE INSULIN PUMP WOULD BE REPLACED. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-722NAP

Patients

Seq Age Sex Outcome Treatment
1 38 YR