FDA Adverse Event
Malfunction
Summary report: N
PARADIGM REAL-TIME INSULIN INFUSION PUMP
MDR report key: 2201042
·
Received July 1, 2011
Report
- Report Number
- 3004209178-2011-82048
- Event Type
- Malfunction
- Date Received
- July 1, 2011
- Date of Event
- June 20, 2011
- Report Date
- June 20, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING FIXED PRIME. TROUBLESHOOTING WAS PERFORMED. RAN A DISPLACEMENT TEST AND THE INSULIN PUMP FAILED THE TEST. ADVISED THE CUSTOMER TO DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO BACK UP PLAN. NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-722NAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 56 YR |