FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2200977 · Received August 11, 2011

Report

Report Number
2531779-2011-05786
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
July 1, 2011
Report Date
July 15, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/04/2011 WITH THE FOLLOWING FINDINGS: THE DOWN ARROW BUTTON WAS FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES AND REQUIRED INCREASED FORCE FOR THE BUTTON TO ENGAGE. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER ALL OF THE BUTTON CONTACTS. UNRELATED TO THE COMPLAINT, THE PUMP DISPLAY SCREEN WAS FOUND TO BE DIM AND DISCOLORED AND THE DISPLAY LENS WAS DETACHED FROM THE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT FOR TWO WEEKS THE DOWN ARROW BUTTON REQUIRED MULTIPLE HARD PRESSES TO ELICIT A RESPONSE. THE PATIENT CONFIRMED THAT THERE WAS NO VISIBLE DAMAGE TO THE KEYPAD. THE PATIENT REPORTEDLY WORE THE PUMP IN AN UNDERGARMENT, OCCASIONALLY USING A PUMP CASE, AND DID NOT CLEAN THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 24 YR