FDA Adverse Event
Malfunction
Summary report: N
ADJ WIRE COLLET 0.7 - 1.8MM
MDR report key: 2200973
·
Received July 1, 2011
Report
- Report Number
- 1811755-2011-02344
- Event Type
- Malfunction
- Date Received
- July 1, 2011
- Date of Event
- June 13, 2011
- Report Date
- June 13, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- KIJ
- PMA / PMN Number
- K972367
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS RETURNED TO THE MFR FOR INVESTIGATION. BASED ON THE QUALITY INVESTIGATION, THE IMPREGLON COATING ON THE DEVICE WAS WORN OFF. THIS CONDITION COULD CAUSE THE DEVICE TO STICK, AND NOT ALLOW THE COLLET TO RELEASE THE WIRE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE WIRE WAS STICKING IN THE DEVICE DURING ROUTINE MAINTENANCE. THERE WAS NO PT INVOLVEMENT AND NO ALLEGATION OF ADVERSE CONSEQUENCES ASSOCIATED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ADJ WIRE COLLET 0.7 - 1.8MM | KIJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |