FDA Adverse Event
Malfunction
Summary report: N
MAESTRO DRILL
MDR report key: 2200959
·
Received July 1, 2011
Report
- Report Number
- 1811755-2011-02308
- Event Type
- Malfunction
- Date Received
- July 1, 2011
- Date of Event
- June 17, 2011
- Report Date
- June 7, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- HBB
- PMA / PMN Number
- K041754
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MD, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE DRILL WAS RETURNED TO THE MANUFACTURER FOR EVALUATION AND UPON INSPECTION, A HOLE WAS DISCOVERED IN THE PNEUMATIC HOSE ASSEMBLY. IT IS UNKNOWN HOW THE HOSE DAMAGE OCCURRED, BUT MAY BE THE RESULT OF MISHANDLING. THE HOSE ASSEMBLY WAS REPLACED AND ADDITIONAL PREVENTATIVE MAINTENANCE WAS ALSO PERFORMED. THE DRILL WAS RETURNED TO THE USER FACILITY.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MANUFACTURER, A HOLE WAS DISCOVERED IN TH HOSE PORTION OF THE PNEUMATIC DRILL. NO OIL LEAKAGE FROM THE DEVICE WAS REPORTED. THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, SO THERE WAS NO PATIENT INVOLVEMENT. NO USER INJURY WAS REPORTED, AND NO OTHER ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MAESTRO DRILL | HBB | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |