FDA Adverse Event Malfunction Summary report: N

CORE MICRO DRILL

MDR report key: 2200950 · Received July 1, 2011

Report

Report Number
1811755-2011-02383
Event Type
Malfunction
Date Received
July 1, 2011
Date of Event
June 8, 2011
Report Date
June 8, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
HBE
PMA / PMN Number
K032303
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

WIDE SPREAD CORROSION DAMAGE WITHIN THE INTERNAL COMPONENTS WAS IDENTIFIED AS THE PROBABLE CAUSE OF THE FAILURE. THE CORROSION CAUSED THE COMPONENTS TO STICK TOGETHER; THIS CAN RESULT IN FRICTION CAUSING EXCESSIVE HEAT PRODUCTION.

Description of Event or Problem · 1

THE CORE MICRO DRILL WAS SENT IN FOR EVALUATION BECAUSE IT WAS OVERHEATING DURING TESTING. THERE WAS NO PATIENT INVOLVEMENT; NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE MICRO DRILL HBE STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK