FDA Adverse Event
Malfunction
Summary report: N
CORE IMPACTION DRILL
MDR report key: 2200946
·
Received July 1, 2011
Report
- Report Number
- 1811755-2011-02382
- Event Type
- Malfunction
- Date Received
- July 1, 2011
- Date of Event
- June 6, 2011
- Report Date
- June 8, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- DZJ
- PMA / PMN Number
- K040369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE DRIVESHAFT AND ROTOR WERE STUCK TOGETHER DUE TO CORROSION. ROTOR RUB WAS IDENTIFIED AS THE PROBABLE CAUSE OF THE OVERHEATING.
Description of Event or Problem · 1
THE CORE IMPACTION DRILL WAS SENT IN FOR EVALUATION DUE TO OVERHEATING DURING A TOOTH EXTRACTION PROCEDURE. THE PROCEDURE WAS COMPLETED WITH A READILY AVAILABLE REPLACEMENT DEVICE WITHOUT ANY PROCEDURE DELAY. NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | CORE IMPACTION DRILL | DZJ | STRYKER INSTRUMENTS KALAMAZOO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |