FDA Adverse Event Injury Summary report: N

STYLE 68 SALINE FILLED BREAST IMPLANT

MDR report key: 22008812 · Received May 13, 2025

Report

Report Number
9617229-2025-07780
Event Type
Injury
Date Received
May 13, 2025
Date of Event
April 14, 2025
Report Date
July 14, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FWM
UDI-DI
10888628000636
PMA / PMN Number
P990074
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: DEFLATION.

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENTS OF DEFLATION WAS RECEIVED ON JULY 15, 2025, WITH LOT NUMBER 3234015. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINT CODES ARE: ¿ DEFLATION: OBSERVED AN OPENING ASSESSED AS FOLD FLAW OPENING. AS PER THE INVESTIGATION PROCEDURE, CREASES AND WEAR ABRASION WERE OBSERVED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 0

ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: B5, D6B, H6.

Description of Event or Problem · 0

PATIENT REPORTED RIGHT SIDE "DEFLATION". HEALTHCARE PROFESSIONAL LATER CONFIRMED. DEVICE REMAINS IMPLANTED.

Description of Event or Problem · 0

PATIENT REPORTED "DEFLATION". THIS RECORD IS FOR THE RIGHT SIDE. DEVICE WAS EXPLANTED AND REPLACED.

Description of Event or Problem · 0

THE DEVICE HAS BEEN EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
2184958 STYLE 68 SALINE FILLED BREAST IMPLANT PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM ALLERGAN (COSTA RICA) 3234015 10888628000636

Patients

Seq Age Sex Outcome Treatment
1 62 YR Female Required Intervention