FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2200867
·
Received August 11, 2011
Report
- Report Number
- 2531779-2011-05779
- Event Type
- Malfunction
- Date Received
- August 11, 2011
- Report Date
- May 23, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS MADE BASED ON THE RESULTS OF EVALUATION COMPLETED (B)(6) 2011. (B)(6). (B)(4). DURING EVALUATION THE FORCE SENSOR ASSEMBLY WAS FOUND TO BE PHYSICALLY DAMAGED AND CONTAMINATION WAS FOUND ON THE FORCE SENSOR. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
Description of Event or Problem · 1
PUMP IS RETURNED FOR MULTIPLE LOSS OF PRIME ALARMS. EVALUATION REVEALED A DAMAGED FORCE SENSOR ASSEMBLY AND CONTAMINATION ON THE FORCE SENSOR. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |