FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2200867 · Received August 11, 2011

Report

Report Number
2531779-2011-05779
Event Type
Malfunction
Date Received
August 11, 2011
Report Date
May 23, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS MADE BASED ON THE RESULTS OF EVALUATION COMPLETED (B)(6) 2011. (B)(6). (B)(4). DURING EVALUATION THE FORCE SENSOR ASSEMBLY WAS FOUND TO BE PHYSICALLY DAMAGED AND CONTAMINATION WAS FOUND ON THE FORCE SENSOR. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

PUMP IS RETURNED FOR MULTIPLE LOSS OF PRIME ALARMS. EVALUATION REVEALED A DAMAGED FORCE SENSOR ASSEMBLY AND CONTAMINATION ON THE FORCE SENSOR. THIS REPORT IS BEING MADE BASED ON THE EVALUATION RESULTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 17 YR