FDA Adverse Event
Injury
Summary report: N
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
MDR report key: 2200791
·
Received July 29, 2011
Report
- Report Number
- 3004209178-2011-82375
- Event Type
- Injury
- Date Received
- July 29, 2011
- Date of Event
- July 16, 2011
- Report Date
- July 18, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE, THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO UNEXPLAINED HIGH BLOOD GLUCOSE. THE BLOOD GLUCOSE READING AT TIME OF CALL WAS 500MG/DL. TROUBLESHOOTING WAS PERFORMED. THE PROGRAMMING ON THE INSULIN PUMP APPEARS TO BE CORRECT. RAN A FIXED PRIME TEST AND THE INSULIN EXITED. PERFORMED A HIGH PRESSURE TEST AND PASSED. NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-723NAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 18 YR | Hospitalization |