FDA Adverse Event
Injury
Summary report: N
SIGN FIN NAIL
MDR report key: 2200771
·
Received July 28, 2011
Report
- Report Number
- 3034525-2011-00030
- Event Type
- Injury
- Date Received
- July 28, 2011
- Date of Event
- July 22, 2011
- Report Date
- July 25, 2011
- Manufacturer
- SURGICAL IMPLANT GENERATION NETWORK (SIGN)
- Product Code
- HSB
- PMA / PMN Number
- K043200
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PK
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
REPORTS OF A BROKEN FIN NAIL AT F/U. REVIEW OF PT X-RAYS SHOWS FIN NAIL BROKEN AT THE PROXIMAL SLOT. SURGEON REPLACED BROKEN FIN NAIL WITH A STANDARD I. M. NAIL. CAUSE OF THE BREAKAGE WAS FULL WEIGHTBEARING ON NONUNION WITH A VERY HIGH FEMUR FRACTURE. NO INDICATION OF PRODUCT DEFECT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SIGN FIN NAIL | SIGN FIN NAIL | HSB | SURGICAL IMPLANT GENERATION NETWORK (SIGN) | F10X280 | 003495 OR 004393 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR | Hospitalization |