FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2200721 · Received August 11, 2011

Report

Report Number
6000001-2011-18174
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
July 20, 2011
Report Date
July 20, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE HAS BEEN RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF MASTER COMPLAINT (B)(4). THE CAUSE WAS DETERMINED TO BE A DAMAGED CORD WRAP. TO CORRECT THE CONDITION, THE CORD WRAP WAS REPLACED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4).THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

DURING PRODUCT EVALUATION BY A BAXTER SERVICE TECHNICIAN, A COLLEAGUE INFUSION PUMP REQUIRED THE REPLACEMENT OF A CORD WRAP. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1