FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2200621 · Received August 11, 2011

Report

Report Number
2649622-2011-11027
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: (B)(4): THE FULL LEAD WAS RETURNED, ANALYZED AND THE OUTER INSULATION WAS BREACHED AND HAD A DEPRESSION. IT WAS NOTED THAT THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THERE WAS BLOOD IN/ON THE HELIX MECHANISM. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE SYSTEM WAS REMOVED DUE TO INFECTION. IT WAS FURTHER REPORTED THAT THE RIGHT ATRIAL LEAD WAS RETURNED, ANALYZED AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 49 YR Hospitalization| R D224TRK IMPLANTABLE PACEMAKER/CARDIO/DEFIB| 4194 IMPLANTABLE PACING LEAD| 6935 IMPLANTABLE TACHY LEAD