ENRHYTHM DR
Report
- Report Number
- 6000144-2011-03601
- Event Type
- Injury
- Date Received
- August 11, 2011
- Manufacturer
- MEDTRONIC MED REL, INC.
- Product Code
- DXY
- PMA / PMN Number
- P980035/S38
- Removal / Correction Number
- Z-1440-2010
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA, US
- Reporter Occupation
- PHYSICIAN
Narratives
THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED AND ANALYZED. CALCULATIONS BASED ON IMPLANT PARAMETERS INDICATE THAT THE DEVICE HAD MET ITS 99.9% EXPECTED LONGEVITY. ANALYSIS OF THE DEVICE MEMORY FOUND THE ERI (ELECTIVE REPLACEMENT INDICATOR) IS THE RESULT OF HIGH INTERNAL BATTERY RESISTANCE. THIS DEVICE IS PART OF THE FIELD ACTION AND HAS TESTED CONSISTENT WITH THE FIELD ACTION.
IT WAS REPORTED THAT THE ATRIAL LEAD HAD AN APPARENT FRACTURE, HIGH IMPEDANCE AND SENSING DIFFICULTY. THE LEAD WAS CAPPED AND REPLACED. THE DEVICE WAS ELECTIVELY EXPLANTED AND REPLACED AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENRHYTHM DR | IMPLANTABLE PULSE GENERATOR | DXY | MEDTRONIC MED REL, INC. | P1501DR | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 83 YR | Hospitalization| R | 5076 IMPLANTABLE PACING LEAD |