FDA Adverse Event Injury Summary report: N

ENRHYTHM DR

MDR report key: 2200601 · Received August 11, 2011

Report

Report Number
6000144-2011-03595
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
DXY
PMA / PMN Number
P980035/S38
Removal / Correction Number
Z-1440-2010
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OK, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS REPORT IS BASED SOLELY ON DEVICE RETURN AND ANALYSIS. NO INFORMATION TO SUGGEST A DEVICE-RELATED ADVERSE EVENT OR PRODUCT PROBLEM WAS RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. THE DEVICE IS PART OF THE ADVISORY FOR THIS MODEL. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED AND ANALYZED. ANALYSIS OF THE DEVICE REVEALED HIGH BATTERY IMPEDANCE. THE DEVICE WAS FULLY FUNCTIONAL, WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Description of Event or Problem · 1

THE DEVICE WAS RETURNED AFTER BEING EXPLANTED DUE TO NORMAL BATTERY DEPLETION. THE DEVICE SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENRHYTHM DR IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC MED REL, INC. P1501DR ASKU

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization| R (B)(4) IMPLANTABLE PACING LEAD