ENRHYTHM MRI
Report
- Report Number
- 6000094-2011-01298
- Event Type
- Injury
- Date Received
- August 11, 2011
- Manufacturer
- MEDTRONIC S.A.
- Product Code
- NVZ
- PMA / PMN Number
- ASKU
- Removal / Correction Number
- ASKU
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. EVALUATION SUMMARY: (B)(4) THE DEVICE WAS RETURNED, ANALYZED AND THE DEVICE WAS FULLY FUNCTIONAL WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.
IT WAS REPORTED THAT IT WAS NOT POSSIBLE TO INTERROGATE THE DEVICE. THE REP TRIED USING TWO DIFFERENT PROGRAMMERS, WITHOUT SUCCESS. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ENRHYTHM MRI | IMPLANTABLE PULSE GENERATOR | NVZ | MEDTRONIC S.A. | EMDR01 | ASKU |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| R |