AUTOMATED IMPELLA CONTROLLER
Report
- Report Number
- 1220648-2025-28328
- Event Type
- Malfunction
- Date Received
- May 12, 2025
- Date of Event
- April 21, 2025
- Report Date
- June 12, 2025
- Manufacturer
- ABIOMED, INC.
- Product Code
- OZD
- UDI-DI
- 00813502011401
- PMA / PMN Number
- P140003
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
THE CONSOLE WAS RETURNED, AND AN EVALUATION IS UNDERWAY. UPON INVESTIGATION CLOSURE, A SUPPLEMENTAL MANUFACTURER DEVICE REPORT WILL BE FILED. THE IMPELLA 5.5 ASSOCIATED WITH THIS AUTOMATED IMPELLA CONTROLLER WAS REPORTED IN MANUFACTURER DEVICE REPORT NUMBER 1220648-2025-28326.
DATA ANALYSIS: ON 3/27 5.5 PUMP 524733 WAS CONNECTED TO (B)(6) AND STARTED RUNNING AT 16:02. 77 MINUTES LATER, THERE IS THE LOG ENTRY "PMD OK STATEREC818 0 2000 257 0 0 0 0XA313 0XA313 CPLD 3" DIRECTLY PRECEDING SAU_MOTOR COMMUNICATION REINITIALIZING AND AN EVENT #115 PUMP STOP ((B)(6) 524733 25_03_27 PUMP STOP 1.PNG). THERE WAS ANOTHER PUMP STOP ON 3/31 WITH SIMILAR LOG ENTRIES BUT NO REINITIALIZATION OF SAU_MOTOR COMMUNICATION ((B)(6) 524733 25_03_31 PUMP STOP 2.PNG). THE PUMP WAS DISCONNECTED AND RECONNECTED ON 4/2 BY THE USER FOR NO DISCERNIBLE REASON IN THE LOGS. ON 4/6, THERE WAS ANOTHER PUMP STOP THAT OCCURRED TWICE IN QUICK SUCCESSION ((B)(6) 524733 25_04_06 PUMP STOP 3.PNG). THE PUMP WAS DISCONNECTED ON 4/7. ON 4/15, A NEW 5.5 PUMP, 581850, WAS CONNECTED TO (B)(6). ON 4/20 THERE IS ANOTHER PUMP STOP ((B)(6) 581850 25_04_20 PUMP STOP 4.PNG) THIS TIME WITH A NEW PUMP. THERE WERE TWO MORE BACK TO BACK PUMP STOPS ON 4/21 AT 8:57 AND THEN AGAIN AT 9:16 ((B)(6) 581850 25_04_21 PUMP STOP 5.PNG). THE PUMP WAS DISCONNECTED ON 4/22. IT WAS ALSO NOTICED FOR EVERY PUMP STOP, THAT AC POWER WAS DISCONNECTED. THE RT LOGS CONFIRM ALL OF THE PUMP STOPS BUT DO NOT EDIFY THE CAUSE. TIME STAMP PUMP AC POWER PMD STATEREC ALARMS PMD RESET AFTER SAU MOTOR ENTRIES SAU MOTOR RESET (B)(6) 2025 17:19:15, 524733. YES OK. 115. YES. YES. YES. (B)(6) 2025 17:19:43, 524733. YES OK. 115+18. NO. NO NO. (B)(6) 2025 13:09:06, 524733. YES OK. 115. YES. NO. NO. (B)(6) 2025 14:35:03, 524733. YES OK. 115. YES. NO. NO. (B)(6) 2025 14:35:06, 524733. YES OK. 115. NO. YES. NO. (B)(6) 2025 14:37:00, 524733. YES OK. 115. YES. YES. NO. (B)(6) 2025 08:38:50, 581850. YES OK. 115. YES. YES. NO. (B)(6) 2025 08:57:04, 581850. YES OK. 115. YES. YES. NO. (B)(6) 2025 09:15:58, 581850. YES OK. 115. YES. YES. YES. DEVICE ANALYSIS: THE CONSOLE WAS RUN OVERNIGHT TWICE WITH A PUMP ON P-8 (ALL PUMP STOPS WERE WITH THE PUMP ON P-8) BUT THE FAILURE MODE WAS NOT REPRODUCED. ADDITIONALLY A PUMP WAS RUN FOR MULTIPLE HOURS ON BATTERY POWER BUT STILL THE FAILURE MODE DID NOT REPRODUCE. ROOT CAUSE: THE CAUSE OF THE MULTIPLE PUMP STOPS ACROSS MULTIPLE PUMPS WAS NOT DETERMINED SINCE THE FAILURE MODE WAS NOT REPRODUCED. ROOT CAUSE: AS A PRECAUTION, PLEASE REPLACE THE IMPELLATRONIC BOARD (0042-1115) AND THE PBM (0042-1125), PLUS CONNECTORS 0042-2709 AND 0042-2711. THEN RUN A TEST PUMP ON BATTERY POWER FOR 30 MINUTES FOLLOWED BY SECTIONS 13-15 OF THE PM (0042-7309) AND FUNCTIONAL CHECK (0042-7316).
IT WAS REPORTED THAT AN IMPELLA 5.5 PUMP WAS IMPLANTED IN A PATIENT BEING ASSESSED FOR TRANSPLANT. THE PATIENT EXPERIENCED A STROKE. THE PATIENT SURVIVED AND LATER THE TEAM OBSERVED A PUMP STOP THAT WAS ABLE TO BE RESTARTED. THE PUMP REMAINS ON FOR SUPPORT. THE ASSOCIATED AUTOMATIC IMPELLA CONTROLLER (AIC) WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1826078 | AUTOMATED IMPELLA CONTROLLER | TEMPORARY NON-ROLLER TYPE LEFT HEART SUPPORT BLOOD PUMP | OZD | ABIOMED, INC. | AUTOMATED IMPELLA CONTROLLER | 1883673 | 00813502011401 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 57 YR | Male | IMPELLA 5.5 |