INSYNC SENTRY

MDR report key: 2200569 · Received August 11, 2011

Report

Report Number
6000144-2011-03597
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
NIK
PMA / PMN Number
P010031/S18
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY (B)(4) ACTUAL LONGEVITY IS < 80% OF 99.9% LONGEVITY LIMIT. THE DEVICE WAS FULLY FUNCTIONAL, WITH NO HIGH CURRENT DRAIN OR EVIDENCE OF BATTERY PROBLEMS. WITHOUT THE HISTORY OF THE PROGRAMMED SETTINGS THROUGHOUT ITS SERVICE LIFE, THERE IS NO WAY TO DETERMINE WHY THE LONGEVITY DID NOT MATCH THE PREDICTED MODEL.

Description of Event or Problem · 1

THE PATIENT'S SPOUSE REPORTED THAT THE LEAD WAS "DEFECTIVE." THE HIGH VOLT PORTION OF THE RIGHT VENTRICULAR LEAD WAS CAPPED. THE PACE/SENSE PORTION IS STILL IN USE. IT WAS ALSO REPORTED THAT THE DEVICE REACHED POSSIBLE PREMATURE ELECTIVE REPLACEMENT INDICATOR. THE DEVICE WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INSYNC SENTRY IMPLANTABLE PACEMAKER/CARDIO/DEFIB NIK MEDTRONIC MED REL, INC. 7299 ASKU

Patients

Seq Age Sex Outcome Treatment
1 69 YR Hospitalization| R 5076 IMPLANTABLE PACING LEAD| 4193 IMPLANTABLE PACING LEAD