FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2200565 · Received August 11, 2011

Report

Report Number
2649622-2011-11013
Event Type
Injury
Date Received
August 11, 2011
Report Date
March 7, 2024
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. ANALYSIS OF THE TWO LEADS ARE IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. EVALUATION SUMMARY: (B)(4): THE DEFIBRILLATOR DEVICE WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. (B)(4): THE ACTUAL LEAD WAS NOT RECEIVED FOR EVALUATION. WE DID RECEIVE PERFORMANCE DATA COLLECTED FROM THE DEVICE AND HAVE ANALYZED THE DATA. DAILY AND WEEKLY IMPEDANCE TRENDS ARE STEADY AND IN RANGE FOR THE RV PACE CIRCUIT. RV LEAD INTEGRITY WARNING ON (B)(6) 2010. 15 SHORT V-V CYCLE EPISODES (<220MS PERIOD) NOTED ON (B)(6) AND (B)(6) 2010. BETWEEN (B)(6) AND (B)(6) 2010 THERE WERE 265 V-SICS (SHORT INTERVAL COUNT) (AVE 34.8 V-SIC/DAY). IN THE DAY PRIOR TO THIS SESSION, THERE WAS ON AVE 111 V-SIC/DAY.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT WAS SHOCKED DUE TO OVERSENSING POSSIBLY CAUSED BY INTERFERENCE OR NOISE. IT WAS LATER REPORTED THAT THE DEVICE HAD REACHED THE END OF SERVICE DUE TO RIGHT VENTRICULAR (RV) LEAD FRACTURES AND MANY SHOCKS. THE RV LEAD AND DEVICE WERE EXPLANTED AND REPLACED. THE ACTIVE ATRIAL LEAD EXHIBITED NOISE AND WAS ALSO EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 22 YR Male Life Threatening| H| R