FDA Adverse Event Injury Summary report: N

CAPSUREFIX NOVUS

MDR report key: 2200560 · Received August 11, 2011

Report

Report Number
2649622-2011-11010
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039/S009
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): PERFORMANCE DATA COLLECTED FROM THE DEVICE WAS ANALYZED, AND REVEALED THAT ON (B)(6) 2010 A PATIENT ALERT OCCURED DUE TO HVB-HVA LEAD IMPEDANCE BEING GREATER THAN 200 OHMS. THE WEEKLY LOG DATA ALSO SHOWED VARYING IMPEDANCE FOR MAX HVB OVER THE RANGE OF 78 TO 240 OHMS BETWEEN (B)(6) 2009 AND (B)(6) 2010. UPON ANALYSIS, NO ANOMALIES FOUND, HOWEVER, THE DEFIBRILLATION COIL WAS DISTORTED, THERE WAS BLOOD/BODY FLUID ON SEVERAL CONDUCTORS (NOT OBSTRUCTED), THE DEFIBRILLATION CONDUCTOR FRACTURED DUE TO OVERSTRESS, THE OUTER INSULATION WAS TORN, THE OUTER INSULATION WAS BREACHED CUT, THERE WAS A WHITE SUBSTANCE ON THE OUTER INSULATION, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS APPARENT EXPLANT DAMAGE AND THE LEAD WAS STRETCHED; PARTIAL LEAD IN SEGMENTS RETURNED AND ANALYZED. (B)(4): NO ANOMALIES FOUND, HOWEVER THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS MELTED, THE OUTER INSULATION HAD COSMETIC ENVIRONMENTAL STRESS CRACKING, THE OUTER INSULATION WAS TORN, THE OUTER INSULATION WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION AND THERE WAS APPARENT EXPLANT DAMAGE; PROXIMAL SEGMENT RETURNED AND ANALYZED. (B)(4): NO ANOMALIES FOUND, HOWEVER THERE WAS BLOOD/BODY FLUID ON THE PROXIMAL CONDUCTOR (NOT OBSTRUCTED), THE OUTER INSULATION WAS BREACHED CUT, THE OUTER INSULATION HAD A COSMETIC DEPRESSION, THERE WAS BLOOD IN/ON THE HELIX/LOBE MECHANISM AND THERE WAS APPARENT EXPLANT DAMAGE; FULL LEAD RETURNED AND ANALYZED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THERE WAS HIGH AND VARYING IMPEDANCE ON THE RIGHT VENTRICULAR LEAD. IT WAS FURTHER REPORTED THAT THE RIGHT VENTRICULAR LEAD WAS EXPLANTED AND REPLACED. DURING THE RIGHT VENTRICULAR LEAD EXTRACTION, THE RIGHT ATRIAL LEAD AND LEFT VENTRICULAR LEAD BOTH BECAME DISLODGED. BOTH OF THESE LEADS WERE ALSO EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX NOVUS IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 5076 ASKU

Patients

Seq Age Sex Outcome Treatment
1 62 YR Hospitalization| R 7299 IMPLANTABLE PACEMAKER/CARDIO/DEFIBRILLATOR