FDA Adverse Event Injury Summary report: N

CAPSUREFIX

MDR report key: 2200515 · Received August 11, 2011

Report

Report Number
2649622-2011-11002
Event Type
Injury
Date Received
August 11, 2011
Manufacturer
MEDTRONIC PUERTO RICO, INC.
Product Code
DTB
PMA / PMN Number
P930039
Removal / Correction Number
ASKU
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. EVALUATION SUMMARY: (B)(4): THE DEVICE WAS RETURNED, ANALYZED AND ANALYSIS OF THE DEVICE REVEALED NORMAL BATTERY DEPLETION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE TRANS-TELEPHONIC MONITORING STRIP SHOWS SPIKES AT 67 BEATS PER MINUTE, NO SENSING AND NO CAPTURE DURING TIMES WHEN THE MAGNET IS NOT ON THE DEVICE. ALSO, NO CLEAR SENSING OR CAPTURE CAN BE SEEN ON NON-MAGNET STRIPS. THE DEVICE WAS EXPLANTED AND REPLACED. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR LEAD HAD LOW IMPEDANCE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CAPSUREFIX IMPLANTABLE PACING LEAD DTB MEDTRONIC PUERTO RICO, INC. 4068 ASKU

Patients

Seq Age Sex Outcome Treatment
1 91 YR Hospitalization| R 5592 IMPLANTABLE PACING LEAD