COLLEAGUE PRE-P1.7
Report
- Report Number
- 6000001-2011-18136
- Event Type
- Malfunction
- Date Received
- August 11, 2011
- Date of Event
- July 10, 2011
- Report Date
- August 4, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE PUMP BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE PUMP WITH A DAMAGED BATTERY WAS CONFIRMED AND REPRODUCED DURING PRODUCT EVALUATION. THE ROOT CAUSE WAS ASSIGNED TO A USE/USER ERROR. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO FIX THE REPORTED CONDITION. ADDITIONAL INFORMATION: THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 6.13.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006.
THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY ALARM. IT IS UNKNOWN WHEN OR IN WHICH CARE AREA THIS EVENT OCCURRED. THIS EVENT MAY HAVE INTERRUPTED DELIVERY. THERE WAS NO REPORT OF PATIENT INVOLVEMENT, INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE-P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |