UNICEL® DXI 800 IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2011-02802
- Event Type
- Malfunction
- Date Received
- August 11, 2011
- Date of Event
- July 11, 2011
- Report Date
- July 11, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K023764
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- OTHER
Narratives
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON-SITE (B)(6), 2011. FSE REPLACED BOTH WASH BUFFER PERI-PUMP TUBING AND THE LIQUID WASTE PERI PUMP TUBING. FSE VERIFIED THE LEAK HAD STOPPED BY PRIMING FLUIDICS AND PERFORMING QC WITHIN THE CUSTOMER'S ESTABLISHED RANGES. FSE DID NOT SPECIFY WHICH PART OR PIECE OF THE TUBING WAS THE CAUSE OF THE LEAK BUT IT WAS REPORTED ONLY WASH BUFFER HAD LEAKED FROM THE INSTRUMENT. A DEFINITIVE ROOT CAUSE FOR THIS EVENT COULD NOT BE DETERMINED. THE BEC INTERNAL IDENTIFIER FOR THIS EVENT IS (B)(4).
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) REGARDING A WASTE LEAK COMING FROM THE LEFT SIDE OF THE UNICEL DXI 800 IMMUNOASSAY SYSTEM. PER CUSTOMER, THE AMOUNT OF LIQUID WAS LESS THAN 1 LITER. THE SPILL WAS CLEANED PER LABORATORY PROCEDURE. NO INJURY OR USER EXPOSURE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXI 800 IMMUNOASSAY SYSTEM | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXI 800 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |