FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 2200397 · Received August 11, 2011

Report

Report Number
6000001-2011-18090
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
July 31, 2011
Report Date
August 5, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K915522
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CUSTOMER?S REPORTED CONDITION OF A FLO-GARD INFUSION PUMP WITH LACKS COVER CLOSED WAS CONFIRMED AND REPRODUCED DURING EVALUATION. THE CAUSE OF THE DETERMINED CONDITION IS A CRACKED/BROKEN DOOR LATCH AND HANDLE. THE DOOR LATCH WAS REPLACED TO FIX THE REPORTED CONDITION.

Additional Manufacturer Narrative · 1

(B)(4).A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE, BUT THE DEVICE HAS NOT YET BEEN RECEIVED BY BAXTER. SHOULD THE DEVICE BE RECEIVED FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A FLO-GARD INFUSION PUMP THAT "LACKS COVER CLOSED"; THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THIS CONDITION WAS FOUND IN THE GENERAL PATIENT WARD UPON POWER UP. THERE WAS NO REPORTED PATIENT INVOLVEMENT, PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1