FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2200390 · Received August 11, 2011

Report

Report Number
6000001-2011-18082
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
July 18, 2011
Report Date
July 19, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K041191
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE PUMP WITH A CRACKED MOTOR COLLAR WAS CONFIRMED DURING PRODUCT EVALUATION. THE ROOT CAUSE WAS ASSIGNED TO A BROKEN TUBE LOAD MOTOR COLLAR. THE MOTOR COLLAR WAS REPLACED TO FIX THE REPORTED CONDITION. THIS INVOLVED AN UNREMEDIATED INFUSION PUMP WITH USER INTERFACE MODULE MASTER SOFTWARE VERSION 5.04.00.

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RECEIVED FOR EVALUATION. UPON COMPLETION OF BAXTER'S INVESTIGATION A FOLLOW UP MEDWATCH WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

THIS IS A REPORT OF A COLLEAGUE INFUSION PUMP WITH CRACKED MOTOR COLLARS, WHICH COULD HAVE CAUSED AN INTERRUPTION OF DELIVERY. IT IS UNKNOWN WHEN THE REPORTED CONDITION OCCURRED, HOWEVER THE REPORTED CONDITION OCCURRED IN THE BIOMEDICAL SERVICE DEPARTMENT. THERE WAS NO PATIENT INVOLVEMENT, INJURY OR MEDICAL INTERVENTION. THE SOFTWARE VERSION FOR THIS DEVICE IS CURRENTLY NOT KNOWN. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1