FDA Adverse Event Injury Summary report: N

SUPERDIMENSION INREACH SYSTEM

MDR report key: 2200363 · Received August 11, 2011

Report

Report Number
3004962788-2011-00039
Event Type
Injury
Date Received
August 11, 2011
Date of Event
June 27, 2011
Report Date
August 10, 2011
Manufacturer
SUPERDIMENSION INC.
Product Code
JAK
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ON (B)(6) 2011 A SITE VISIT WAS PERFORMED. AN ACCURACY TEST AND FUNCTIONAL TEST WERE BOTH PERFORMED AND BOTH PASSED. DURING THE VISIT STAFF MEMBERS WERE INQUIRING ABOUT DIFFERENT BED POSITIONS AND HOW THEY MIGHT AFFECT ACCURACY DURING CASES. THEY NOTED THAT DIFFERENT BED POSITIONS HAD BEEN USED DURING RECENT PROCEDURES. THE SUPERDIMENSION TECHNICAL SPECIALIST EXPLAINED THAT THE BED MUST BE IN THE SAME LOCATION IT WAS MAPPED IN FOR EVERY CASE AND CANNOT BE MOVED. IN ADDITION, HE FURTHER EXPLAINED THAT THIS CAN AFFECT THE ACCURACY OF THE SYSTEM AS DESCRIBED IN THE USER MANUAL. THE SITE HAS PERFORMED TWO SUCCESSFUL CASES SINCE THE SITE VISIT WITH NO ISSUES OR COMPLICATIONS REPORTED. PNEUMOTHORAX IS A KNOWN COMPLICATION WHEN A LUNG BIOPSY IS PERFORMED DURING A TRANSBRONCHIAL LUNG BIOPSY, OR CT-GUIDED PERCUTANEOUS BIOPSY WITH COMPLICATION RATES UP TO APPROXIMATELY 27% WITH THE CT GUIDED PROCEDURE. BIOPSY TOOLS ARE DESIGNED TO HAVE SHARP EDGES AND THEIR FUNCTION IS TO PUNCTURE OR OTHERWISE PENETRATE THE TISSUE IN ORDER TO COLLECT A SPECIMEN. THE USER MANUAL STATES THE FOLLOWING: CAUTION: THERE ARE CERTAIN DETAILS THAT MUST BE VERIFIED BEFORE THE DAILY SETUP IN ORDER TO ASSURE LOCATION ACCURACY. SPECIFICALLY: THE BED ARRANGEMENT (FOR EXAMPLE, THE TABLE-TOP POSITION WITH RESPECT TO SUPPORT LEG AND RAIL POSITION) MUST BE IDENTICAL TO ITS ARRANGEMENT DURING SYSTEM SETUP. THE PROCEDURE ROOM SHOULD BE ARRANGED ACCORDING TO THE PROCEDURE ROOM CONFIGURATION FORM PROVIDED BY THE SUPERDIMENSION TECHNICIAN FOLLOWING THE INSTALLATION OF THE SYSTEM, AND AS APPEARS IN THE PROCEDURE ROOM CONFIGURATION PICTURES.

Description of Event or Problem · 1

THE SITE REPORTED THAT THEY HAD A CASE IN (B)(6) THAT HAD RESULTED IN A PNEUMOTHORAX. IT WAS REPORTED THAT ON (B)(6) 2011, THE PHYSICIAN PERFORMED A SUPERDIMENSION CASE UNDER CONSCIOUS SEDATION AND HAD BIOPSIED ON BOTH SIDES OF THE PATIENTS LUNGS. THERE WAS NOTHING NOTABLE FROM THE CASE REPORTED. THE PNEUMOTHORAX WAS IDENTIFIED LATER THAT DAY AND REQUIRED A CHEST TUBE. IT WAS REPORTED THAT THE PATIENT WAS DISCHARGED ON (B)(6) 2011 AND NO FURTHER COMPLICATIONS WERE REPORTED. THE PHYSICIAN INFORMED SUPERDIMENSION OF THE PNEUMOTHORAX AND WAS CONCERNED THE SYSTEM ACCURACY NEEDED TO BE CHECKED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SUPERDIMENSION INREACH SYSTEM ELECTROMAGNETIC NAVIGATION BRONCHOSCOPY (ENB) JAK SUPERDIMENSION INC.

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization