FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2200324 · Received August 11, 2011

Report

Report Number
6000001-2011-18029
Event Type
Malfunction
Date Received
August 11, 2011
Date of Event
May 26, 2011
Report Date
July 20, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RECEIVED BY BAXTER, AND THE CONDITION WAS CONFIRMED BY SERVICE. THIS COMPLAINT IS AN ANCILLARY SERVICE EVENT OPENED DURING INVESTIGATION OF (B)(4). THE CAUSE WAS DETERMINED TO BE FAULTY MAIN BATTERIES. TO CORRECT THIS CONDITION, THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP MDR WILL BE SENT.

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Description of Event or Problem · 1

DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER PERSONNEL, A COLLEAGUE INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED A BATTERY DEPLETED SET ALARM, WHICH INTERRUPTED DELIVERY. THIS ALARM MAY HAVE BEEN A FALSE ALARM. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90. NO ADDITIONAL INFORMATION IS AVAILABLE.