COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-18021
- Event Type
- Malfunction
- Date Received
- August 11, 2011
- Date of Event
- April 26, 2011
- Report Date
- July 20, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS REPAIRED ON-SITE AT THE CUSTOMER FACILITY BY A BAXTER FIELD SERVICE TECHNICIAN, AND THE CONDITION WAS CONFIRMED. THIS COMPLAINT IS AN ANCILLARY OF (B)(4). THE CAUSE WAS NOT IDENTIFIED AS THE CONDITION WAS NOT RELATED TO THE CUSTOMER REPORTED CONDITION. NO FURTHER TESTING HAS BEEN COMPLETED AT THIS TIME AND NO REPAIRS HAVE BEEN PERFORMED. IF ANY ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW-UP MDR WILL BE SENT.
(B)(4). THE DEVICE WAS INVESTIGATED BY A BAXTER FIELD SERVICE ENGINEER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
DURING A REVIEW OF THE PUMP'S EVENT HISTORY BY BAXTER PERSONNEL, A COLLEAGUE INFUSION PUMP WAS FOUND TO HAVE EXPERIENCED A DOWNSTREAM OCCLUSION ALARM, WHICH INTERRUPTED DELIVERY. THIS ALARM MAY HAVE BEEN A FALSE ALARM. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THIS EVENT. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |