SUPERDIMENSION INREACH SYSTEM
Report
- Report Number
- 3004962788-2011-00038
- Event Type
- Injury
- Date Received
- August 11, 2011
- Date of Event
- June 24, 2011
- Report Date
- August 10, 2011
- Manufacturer
- SUPERDIMENSION INC.
- Product Code
- JAK
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). ON (B)(6) 2011 A SITE VISIT WAS PERFORMED. AN ACCURACY TEST AND FUNCTIONAL TEST WERE BOTH PERFORMED AND BOTH PASSED. DURING THE VISIT STAFF MEMBERS WERE INQUIRING ABOUT DIFFERENT BED POSITIONS AND HOW THEY MIGHT AFFECT ACCURACY DURING CASES. THEY NOTED THAT DIFFERENT BED POSITIONS HAD BEEN USED DURING RECENT PROCEDURES. THE SUPERDIMENSION TECHNICAL SPECIALIST EXPLAINED THAT THE BED MUST BE IN THE SAME LOCATION IT WAS MAPPED IN FOR EVERY CASE AND CANNOT BE MOVED. IN ADDITION, HE FURTHER EXPLAINED THAT THIS CAN AFFECT THE ACCURACY OF THE SYSTEM AS DESCRIBED IN THE USER MANUAL. PNEUMOTHORAX IS A KNOWN COMPLICATION WHEN A LUNG BIOPSY IS PERFORMED DURING A TRANSBRONCHIAL LUNG BIOPSY, OR CT-GUIDED PERCUTANEOUS BIOPSY WITH COMPLICATION RATES UP TO APPROXIMATELY 27% WITH THE CT GUIDED PROCEDURE. BIOPSY TOOLS ARE DESIGNED TO HAVE SHARP EDGES AND THEIR FUNCTION IS TO PUNCTURE OR OTHERWISE PENETRATE THE TISSUE IN ORDER TO COLLECT A SPECIMEN. THE USER MANUAL STATES THE FOLLOWING: CAUTION: THERE ARE CERTAIN DETAILS THAT MUST BE VERIFIED BEFORE THE DAILY SETUP IN ORDER TO ASSURE LOCATION ACCURACY. SPECIFICALLY: THE BED ARRANGEMENT (FOR EXAMPLE, THE TABLE-TOP POSITION WITH RESPECT TO SUPPORT LEG AND RAIL POSITION) MUST BE IDENTICAL TO ITS ARRANGEMENT DURING SYSTEM SETUP. THE PROCEDURE ROOM SHOULD BE ARRANGED ACCORDING TO THE PROCEDURE ROOM CONFIGURATION FORM PROVIDED BY THE SUPERDIMENSION TECHNICIAN FOLLOWING THE INSTALLATION OF THE SYSTEM, AND AS APPEARS IN THE PROCEDURE ROOM CONFIGURATION PICTURES.
THE SITE REPORTED THAT THEY HAD A CASE IN (B)(6) THAT HAD RESULTED IN A PNEUMOTHORAX. IT WAS REPORTED THAT ON (B)(6) 2011, THE PHYSICIAN FOUND A PNEUMOTHORAX POST PROCEDURE. THE CASE HAD BEEN PERFORMED UNDER CONSCIOUS SEDATION. THERE WAS NOTHING NOTABLE FROM THE CASE REPORTED. THE PHYSICIAN INFORMED SUPERDIMENSION OF THE PNEUMOTHORAX AND WAS CONCERNED THE SYSTEM ACCURACY NEEDED TO BE CHECKED. THE PATIENT WAS OBSERVED AND DISCHARGED LATER THAT SAME DAY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SUPERDIMENSION INREACH SYSTEM | ELECTROMAGNETIC NAVIGATION BRONCHOSCOPY (ENB) | JAK | SUPERDIMENSION INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |