COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-18012
- Event Type
- Malfunction
- Date Received
- August 11, 2011
- Date of Event
- July 25, 2011
- Report Date
- July 26, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF A DAMAGED BATTERY WAS CONFIRMED AND REPRODUCED DURING ON-SITE PRODUCT EVALUATION. THE ROOT CAUSE WAS IDENTIFIED AS DEPLETED MAIN BATTERIES AS A RESULT OF USE ERROR. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT REPORTED CONDITION. THE USER INTERFACE MODULE SOFTWARE VERSION IS 6.13.90. A SERVICE HISTORY REVIEW WAS PERFORMED REVEALING THE REPORTED CONDITION IS NOT RELATED TO ANY PREVIOUS CUSTOMER SERVICE REQUEST ON THIS PUMP. HOWEVER, DURING PREVIOUS SERVICE THE BATTERIES AND HARNESS WERE REPLACED.
(B)(4). THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. THE USER INTERFACE SOFTWARE VERSION IS CURRENTLY UNKNOWN. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |