Additional Manufacturer Narrative · 1

(B)(4). AS THE DATE OF ONSET OF THIS PERITONITIS EPISODE IS UNKNOWN AND PATIENTS DISCARD SUPPLIES AFTER EACH THERAPY, THE SAMPLE WAS NOT REQUESTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. THIS IS REPORT 4 OF 5 INVOLVED IN THIS PERITONITIS EVENT.

Additional Manufacturer Narrative · 1

(B)(4). A BATCH REVIEW WAS PERFORMED FOR POTENTIALLY ASSOCIATED LOT NUMBERS GD885301, GD884437, AND GD883520 WITH NO DEFECTS NOTED. THE CAUSE OF THE PERITONITIS WAS UNDETERMINED. BAXTER HAS RECEIVED SIMILAR REPORTS FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH CAPA (B)(4).

Description of Event or Problem · 1

INITIALLY, THE PATIENT CALLED THE BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) REGARDING A CHECK SUPPLY LINE ALARM, WHICH OCCURRED ON THE HOMECHOICE (HC), DURING PRIME. THE TSR ASSISTED THE HP TO COMPLETE THE PRIME. THE SOLUTION WAS PROVIDED OVER THE PHONE AND THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE INITIAL CALL. ON (B)(4) 2011, DURING A FOLLOW UP CALL WITH THE PATIENT BY BAXTER PRODUCT SURVEILLANCE REGARDING THE CHECK SUPPLY LINE ALARM, THE PATIENT STATED THE SAMPLE HAD BEEN DISCARDED AND THERE WAS NO LOT NUMBER AVAILABLE. DURING THE CONVERSATION, THE HP SAID THE NEXT DAY AFTER THE INCIDENT WITH THE ALARM, HE WENT TO HIS CLINIC AND HE WAS DIAGNOSED WITH PERITONITIS. THE HP SAID HE WAS GIVEN ANTIBIOTICS AND THAT HE IS FEELING BETTER. NO FURTHER INFORMATION WAS PROVIDED BY THE CONSUMER ON THE INITIAL FOLLOW UP CALL. ON (B)(4) 2011, PRODUCT SURVEILLANCE FOLLOWED UP WITH THE PATIENT'S PERITONEAL DIALYSIS NURSE (PDN) WHO PROVIDED THE FOLLOWING INFORMATION. THE PDN CONFIRMED THAT THE PATIENT WAS DIAGNOSED WITH PERITONITIS COINCIDENT WITH 2.5 % DIANEAL AMBUFLEX AND ICODEXTRIN THERAPY (DOSES, FREQUENCIES AND LOTS WERE NOT REPORTED). THE PATIENT WAS NOT HOSPITALIZED AND RECEIVED REMEDIAL TREATMENT WITH AN UNSPECIFIED ANTIBIOTIC. ON AN UNSPECIFIED DATE THE PATIENT RECOVERED FROM THE PERITONITIS. CONCOMITANT MEDICATIONS WERE NOT REPORTED. THE PDN STATED THE PATIENT IS USING UNSPECIFIED DISPOSABLE PRODUCTS WITH LUER CONNECTORS. THE PDN STATED THAT THE PATIENT WAS EXPERIENCING DIARRHEA AT THE TIME THAT THE PERITONITIS OCCURRED. THE PDN STATED THE CAUSE OF THE PERITONITIS WAS UNKNOWN, HOWEVER, COULD BE INTERNAL. THE PDN DID NOT CONSIDER THE PERITONITIS TO BE RELATED TO THE PD THERAPY AND NOT RELATED TO ANY BAXTER PRODUCT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MINICAP DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM KDI BAXTER HEALTHCARE - CLEVELAND

Patients

Seq Age Sex Outcome Treatment
1 52 YR Required Intervention ICODEXTRIN| 2.5 % DIANEAL AMBUFLEX