FDA Adverse Event Injury Summary report: N

STOCKERT 70 RF GENERATOR

MDR report key: 2200189 · Received August 10, 2011

Report

Report Number
9612355-2011-00024
Event Type
Injury
Date Received
August 10, 2011
Date of Event
July 7, 2011
Report Date
July 12, 2011
Manufacturer
STOCKERT GMBH
Product Code
DRF
PMA / PMN Number
P990071
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE FIELD SERVICE ENGINEER SPOKE TO CUSTOMER AND THEY STATED THAT THE ISSUE WAS NOT STOCKERT GENERATOR RELATED. THEY MOVED THE STOCKERT GENERATOR TO A NEW LAB ON THE SAME DAY AND IT PERFORMED NORMALLY. IT WAS MOST LIKELY A PATCH PLACEMENT WORK FLOW ISSUE. CUSTOMER DECLINED REPAIR. THE DEVICE HISTORY RECORD FOR STOCKERT GENERATOR (B)(4) SHOWS THAT NO MANUFACTURING OR TEST FAILS WERE NOTED DURING THE MANUFACTURING CYCLE RELATED TO FUNCTIONALITY OF THE DEVICE. THE DEVICE MET ALL REQUIREMENTS PRIOR TO DISTRIBUTION. THE CUSTOMER IS CURRENTLY USING THIS DEVICE WITHOUT ANY ISSUES.

Additional Manufacturer Narrative · 1

CONCOMITANT PRODUCT: EZ STEER; THERMOCOOL BI-DIRECTIONAL CATHETER, CATALOG # BDI75TCFJRT, LOT # UNKNOWN. (B)(4).

Description of Event or Problem · 1

IN AN AFIB ABLATION PROCEDURE, IT WAS REPORTED THAT THERE WERE HIGH IMPEDANCE READINGS ON THE STOCKERT WHILE IN USE WITH ESI MAPPING SYSTEM (NON-BWI PRODUCT) AND A VALLEY LAB PATCH. THE IMPEDANCE READING WAS REPORTED AT 230 OHMS. THERE WAS A BURN TO THE PATIENT'S LOWER LEFT FLANK BETWEEN THE BUTTOCKS AND THE MID BACK APPROXIMATELY THE SIZE OF A SILVER DOLLAR. THE PATIENT IS GOING TO A WOUND CARE CENTER TODAY FOR TREATMENT TO THE BURN. THERE WERE NO ERROR MESSAGES TO INDICATE THAT THERE WAS ANY SYSTEM MALFUNCTION. THE INDIFFERENT ELECTRODE CABLE AND PATCH WERE MOVED FROM THE BELLY TO THE LOWER LEG TO TRY TO REDUCE THE IMPEDANCE BUT THIS WAS UNSUCCESSFUL. THE CATHETER WAS CHANGED OUT AND THIS ALSO DID NOT RESOLVE THE IMPEDANCE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STOCKERT 70 RF GENERATOR RADIOFREQUENCY GENERATOR AND ACCESSORIES/ACCESSORY CABLE DRF STOCKERT GMBH M-5463-01

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention