STOCKERT 70 RF GENERATOR
Report
- Report Number
- 9612355-2011-00024
- Event Type
- Injury
- Date Received
- August 10, 2011
- Date of Event
- July 7, 2011
- Report Date
- July 12, 2011
- Manufacturer
- STOCKERT GMBH
- Product Code
- DRF
- PMA / PMN Number
- P990071
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE FIELD SERVICE ENGINEER SPOKE TO CUSTOMER AND THEY STATED THAT THE ISSUE WAS NOT STOCKERT GENERATOR RELATED. THEY MOVED THE STOCKERT GENERATOR TO A NEW LAB ON THE SAME DAY AND IT PERFORMED NORMALLY. IT WAS MOST LIKELY A PATCH PLACEMENT WORK FLOW ISSUE. CUSTOMER DECLINED REPAIR. THE DEVICE HISTORY RECORD FOR STOCKERT GENERATOR (B)(4) SHOWS THAT NO MANUFACTURING OR TEST FAILS WERE NOTED DURING THE MANUFACTURING CYCLE RELATED TO FUNCTIONALITY OF THE DEVICE. THE DEVICE MET ALL REQUIREMENTS PRIOR TO DISTRIBUTION. THE CUSTOMER IS CURRENTLY USING THIS DEVICE WITHOUT ANY ISSUES.
CONCOMITANT PRODUCT: EZ STEER; THERMOCOOL BI-DIRECTIONAL CATHETER, CATALOG # BDI75TCFJRT, LOT # UNKNOWN. (B)(4).
IN AN AFIB ABLATION PROCEDURE, IT WAS REPORTED THAT THERE WERE HIGH IMPEDANCE READINGS ON THE STOCKERT WHILE IN USE WITH ESI MAPPING SYSTEM (NON-BWI PRODUCT) AND A VALLEY LAB PATCH. THE IMPEDANCE READING WAS REPORTED AT 230 OHMS. THERE WAS A BURN TO THE PATIENT'S LOWER LEFT FLANK BETWEEN THE BUTTOCKS AND THE MID BACK APPROXIMATELY THE SIZE OF A SILVER DOLLAR. THE PATIENT IS GOING TO A WOUND CARE CENTER TODAY FOR TREATMENT TO THE BURN. THERE WERE NO ERROR MESSAGES TO INDICATE THAT THERE WAS ANY SYSTEM MALFUNCTION. THE INDIFFERENT ELECTRODE CABLE AND PATCH WERE MOVED FROM THE BELLY TO THE LOWER LEG TO TRY TO REDUCE THE IMPEDANCE BUT THIS WAS UNSUCCESSFUL. THE CATHETER WAS CHANGED OUT AND THIS ALSO DID NOT RESOLVE THE IMPEDANCE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | STOCKERT 70 RF GENERATOR | RADIOFREQUENCY GENERATOR AND ACCESSORIES/ACCESSORY CABLE | DRF | STOCKERT GMBH | M-5463-01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |