FDA Adverse Event Death Summary report: N

ENPULSE

MDR report key: 2200173 · Received August 10, 2011

Report

Report Number
6000144-2011-03582
Event Type
Death
Date Received
August 10, 2011
Date of Event
August 1, 2005
Manufacturer
MEDTRONIC MED REL, INC.
Product Code
DXY
PMA / PMN Number
P980035/S032
Removal / Correction Number
ASKU
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: (B)(4): THE DEVICE WAS RETURNED AND ANALYZED AND FOUND TO HAVE NORMAL BATTERY DEPLETION. THE DEVICE MET EXPECTED LONGEVITY. (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. THE DEVICE(S) ASSOCIATED WITH THIS ADVERSE OUTCOME WAS/WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. CONSEQUENTLY, CONTACT INFORMATION TO COMPLETE FOLLOW-UP IS NOT REASONABLY KNOWN. THEREFORE, ATTEMPTS FOR ADDITIONAL INFORMATION CANNOT BE MADE. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

EVALUATION SUMMARY: (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. (B)(4): THE PROXIMAL SEGMENT OF THE LEAD WAS RETURNED, ANALYZED AND NO ANOMALIES WERE FOUND. THE DEVICE(S) ASSOCIATED WITH THIS ADVERSE OUTCOME WAS/WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. CONSEQUENTLY, CONTACT INFORMATION TO COMPLETE FOLLOW-UP IS NOT REASONABLY KNOWN. THEREFORE, ATTEMPTS FOR ADDITIONAL INFORMATION CANNOT BE MADE. ANALYSIS OF THE DEVICE ISI IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

THE DEVICE(S) ASSOCIATED WITH THIS ADVERSE OUTCOME WAS/WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. CONSEQUENTLY, CONTACT INFORMATION TO COMPLETE FOLLOW-UP IS NOT REASONABLY KNOWN. THEREFORE, ATTEMPTS FOR ADDITIONAL INFORMATION CANNOT BE MADE. ANALYSIS OF THE DEVICE AND LEADS ARE IN PROCESS; THE RESULTS WILL BE FORWARDED WHEN AVAILABLE. THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

AN IMPLANTABLE PULSE GENERATOR (IPG) AND TWO LEADS WERE RETURNED FROM AN UNKNOWN SOURCE WITH NO INFORMATION. INFORMATION IDENTIFIED IN THE MANUFACTURE'S DATA BASE INDICATED THE PATIENT DIED APPROXIMATELY FOUR MONTHS POST THE IMPLANT OF THE IPG SYSTEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ENPULSE IMPLANTABLE PULSE GENERATOR DXY MEDTRONIC MED REL, INC. E2DR21 ASKU

Patients

Seq Age Sex Outcome Treatment
1 85 YR Death