ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Report
- Report Number
- 1030489-2011-01021
- Event Type
- Malfunction
- Date Received
- August 10, 2011
- Report Date
- July 13, 2011
- Manufacturer
- MEDTRONIC SOFAMOR DANEK INSTRUMENT MANUFACTURING
- Product Code
- LXH
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR EVALUATION. THE SUPERIOR SHAFT IS BROKEN OFF AT THE CENTER OF THE JAW PIVOT PIN FORWARD. THE BROKEN OFF PORTION OF THE SHAFT IS MISSING AND WAS NOT RETURNED FOR ANALYSIS. OPTICAL EXAMINATION IDENTIFIED A FAIRLY BRITTLE FRACTURE, WITH NO INDICATION OF FATIGUE. THE LOCATION, DIRECTION, AND AMOUNT OF FORCE REQUIRED IN ORDER INDUCE FRACTURE OF THE SHAFT IS CONSISTENT WITH APPLICATION OF EXCESSIVE FORCE, RESULTING IN THE FOREGOING EVENT. A REVIEW OF THE DEVICE HISTORY RECORDS FOR THIS DEVICE DID NOT REVEAL ANY NON-CONFORMANCES TO SPECIFICATION OR DEVIATIONS IN PROCEDURE WHICH MIGHT CONTRIBUTE TO THE REPORTED EVENT.
IT WAS REPORTED THAT THE PATIENT UNDERWENT A SPINAL PROCEDURE AND THE TIP OF THE INSTRUMENT CRACKED. NO PATIENT COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | ORTHOPEDIC MANUAL SURGICAL INSTRUMENT | LXH | MEDTRONIC SOFAMOR DANEK INSTRUMENT MANUFACTURING | NA | GZ05G041 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |