CD HORIZON® SPINAL SYSTEM
Report
- Report Number
- 1030489-2025-01944
- Event Type
- Malfunction
- Date Received
- May 12, 2025
- Date of Event
- February 7, 2025
- Report Date
- May 12, 2025
- Manufacturer
- MEDTRONIC SOFAMOR DANEK USA, INC
- Product Code
- HXX
- UDI-DI
- 00643169043640
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- 003
Narratives
H3: PRODUCT ANALYSIS FOR PART # 5584111; LOT # K23D1225 VISUAL INSPECTION CONFIRMED THE TORX TIP OF INSTRUMENT HAS BEEN DAMAGED AND THE ATTACHMENT PIN TO THE INTERNAL BUSHING HAS BROKEN. OPTICAL EXAMINATION REVEALED THE TORX TIP EDGES HAVE BEEN ROLLED OVER AND DEFORMED. THIS TYPE OF DAMAGE IS CONSISTENT WITH TORSIONAL OVERLOAD. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
INFORMATION WAS RECEIVED FROM USER FACILITY VIA A MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR OLIF / PSF SPINAL THERAPY. IT WAS REPORTED THAT THE INSERTER IS LOOSE. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 347500 | CD HORIZON® SPINAL SYSTEM | SCREWDRIVER | HXX | MEDTRONIC SOFAMOR DANEK USA, INC | 5584111 | K23D1225 | 00643169043640 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |