Narratives
Additional Manufacturer Narrative · 0
(B)(4).
Additional Manufacturer Narrative · 0
(B)(4). 3004753838-2025-113546 WAS REPORTED IN ERROR. PLEASE DISREGARD INITIAL REPORTING OF THIS EVENT AS THIS EVENT HAS NOW BEEN DEEMED NOT REPORTABLE.
Description of Event or Problem · 0
IT WAS REPORTED THAT NO READINGS/ SENSOR ERROR/ BRIEF SENSOR ISSUE OCCURRED. PERFORMANCE DATA WAS REVIEWED. NO READINGS/ SENSOR ERROR/ BRIEF SENSOR ISSUE WAS NOT FOUND WITHIN THE INVESTIGATION WINDOW. THE ALLEGATION WAS NOT CONFIRMED. HOWEVER, SIGNAL LOSS OVER ONE HOUR WAS FOUND IN CONNECTION TO THE REPORTED ALLEGATION. THE PROBABLE CAUSE OF THE SIGNAL LOSS COULD NOT BE DETERMINED. NO INJURY OR MEDICAL INTERVENTION WAS REPORTED.
Description of Event or Problem · 0
AFTER SUBMISSION OF THE INITIAL MDR, PRODUCT WAS RECEIVED ON 5/21/2025 AND IT WAS DETERMINED THIS COMPLAINT IS NOT REPORTABLE PER (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1230604 | DEXCOM G7 CONTINUOUS GLUCOSE MONITORING SYSTEM | CONTINUOUS GLUCOSE MONITOR | QBJ | DEXCOM, INC. | 9500-161 | 1825058001 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Female | MEDRONIC PUMP |