FDA Adverse Event Injury Summary report: N

SMR SHOULDER

MDR report key: 21993924 · Received May 9, 2025

Report

Report Number
3008021110-2025-00054
Event Type
Injury
Date Received
May 9, 2025
Date of Event
March 28, 2025
Report Date
May 9, 2025
Manufacturer
LIMACORPORATE S.P.A.
Product Code
HSD
UDI-DI
08033390002028
PMA / PMN Number
K100858
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

CHECKING THE MANUFACTURING CHARTS OF THE INVOLVED LOT #1601336, NO PRE-EXISTING ANOMALY WAS FOUND ON THE 3 DEVICES MANUFACTURED WITH THE SAME LOT # - STER. THIS IS THE FIRST AND ONLY COMPLAINT RECEIVED ON THAT LOT #. WE SUBMIT A FINAL MDR WHEN THE INVESTIGATION IS COMPLETE.

Description of Event or Problem · 0

SHOULDER REVISION SURGERY OF A SMR ANATOMIC HEMI PROSTHESIS PERFORMED ON (B)(6)2025, DUE TO POOR BONE QUALITY. THIS IS A SECOND STAGE REVISION SURGERY: IT WAS REPORTED THAT THE PATIENT HAD A PRIOR ANATOMIC SHOULDER WITH A CEMENTED 3-PEGGED POLY GLENOID (PRODUCT CODE AND LOT # UNKNOWN). THE PATIENT WAS REVISED AND THE CEMENTED POLY GLENOID WAS REMOVED (DATE OF THE SURGERY UNKNOWN). THE GLENOID WAS PACKED WITH A BONE GRAFT AND LEFT TO HEAL FOR THE SECOND STAGE REVISION. AT THE TIME, AN ANATOMIC HUMERAL BODY AND A HUMERAL HEAD WERE IMPLANTED, AND THE PATIENT WAS LEFT WITH THE HEMI PROSTHESIS. UPON INSPECTION OF THE GLENOID ON THE SURGERY OF (B)(6) 2025, THE BONE QUALITY WAS DEEMED TOO POOR TO PROCEED TO A REVERSE BASEPLATE. THE FOLLOWING DEVICES WERE EXPLANTED: · SMR TRAUMA HUMERAL BODY # SHORT (PRODUCT CODE 1350.15.030, LOT #1601336 - STER. (B)(4)). · SMR HUMERAL HEAD D.44 H.14MM (PRODUCT CODE 1321.09.441, LOT #2014027 - STER. (B)(4)). · NEUTRAL ADAPTOR TAPER STANDARD (PRODUCT CODE 1330.15.270, LOT #2014099 - STER. (B)(4)). THE SAME SIZE COMPONENTS WERE PUT BACK IN AND THE PATIENT WILL HAVE A HEMI FOR THE FORESEEABLE FUTURE. PATIENT IS A FEMALE, 81 YEARS OLD. POOR BONE QUALITY REPORTED. EVENT HAPPENED IN THE US.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1034407 SMR SHOULDER TRAUMA HUMERAL BODIES WITH LOCKING SCREW SHORT HSD LIMACORPORATE S.P.A. HUMERAL BODY 1601336 08033390002028

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention