FDA Adverse Event Malfunction Summary report: N

ALINITY I PROCESSING MODULE

MDR report key: 21984002 · Received May 8, 2025

Report

Report Number
3016438761-2025-00270
Event Type
Malfunction
Date Received
May 8, 2025
Date of Event
April 19, 2025
Report Date
June 13, 2025
Manufacturer
ABBOTT LABORATORIES
Product Code
JJE
UDI-DI
00380740137366
PMA / PMN Number
K170317
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BG
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AFTER FURTHER EVALUATION, THE SUSPECT MEDICAL DEVICE WAS CHANGED FROM THE ALINITY I PROCESSING MODULE, LIST 03R65-01, ABBOTT LABORATORIES, IRVING, TEXAS, USA TO THE ALINITY I FREE T4, LIST 07P70-30, ABBOTT IRELAND DIAGNOSTICS DIVISION, LONGFORD, IRELAND. MDR NUMBER 3005094123-2025-00295 HAS BEEN SUBMITTED AND ALL FURTHER INFORMATION WILL BE DOCUMENTED UNDER THAT MDR NUMBER.

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I FREE T4 RESULT GENERATED BY THE ALINITY I PROCESSING MODULE FOR A PATIENT. THE SAMPLE WAS REPEATED, AND A LOWER RESULT WAS OBTAINED. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6): INITIAL RESULT = 2.65 NG/DL; REPEAT RESULT = 1.08 NG/DL THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Description of Event or Problem · 0

THE CUSTOMER OBSERVED FALSELY ELEVATED ALINITY I FREE T4 RESULT GENERATED BY THE ALINITY I PROCESSING MODULE FOR A PATIENT. THE SAMPLE WAS REPEATED, AND A LOWER RESULT WAS OBTAINED. THE FOLLOWING DATA WAS PROVIDED: SID (B)(6): INITIAL RESULT = 2.65 NG/DL; REPEAT RESULT = 1.08 NG/DL THERE WAS NO IMPACT TO PATIENT MANAGEMENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1103150 ALINITY I PROCESSING MODULE ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE ABBOTT LABORATORIES 00380740137366

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown ALNTY I FREE T4 1200T, 07P70-30, (B)(6)| ALNTY I FREE T4 1200T, 07P70-30, (B)(6)