FDA Adverse Event
Malfunction
Summary report: N
QUANTUM WORKSTATION 12" ELITE
MDR report key: 21979600
·
Received May 7, 2025
Report
- Report Number
- 3006073153-2025-00114
- Event Type
- Malfunction
- Date Received
- May 7, 2025
- Date of Event
- April 4, 2025
- Report Date
- May 7, 2025
- Manufacturer
- SPECTRUM MEDICAL LTD
- Product Code
- DRY
- UDI-DI
- 05060434422309
- PMA / PMN Number
- K202557
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
Additional Manufacturer Narrative · 0
INVESTIGATION FOUND THAT USER UNINTENTIONALLY SHUT DOWN THE SYSTEM DURING THE PROCEDURE. SPECTRUM MEDICAL LTD HAS CONSIDERED THIS POTENTIAL USE ERROR IN IMPROVEMENTS TO THE SOFTWARE, WHICH PREVENTS SHUTDOWN COMMANDS WHILE THE PUMP IS ACTIVE.
Description of Event or Problem · 0
USER REPORTED SYSTEM SCREENS WENT BLANK AND PUMP STOPPED. THERE WAS NO PATIENT HARM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 144421 | QUANTUM WORKSTATION 12" ELITE | MONITOR, BLOOD-GAS, ON-LINE, CARDIOPULMONARY BYPASS | DRY | SPECTRUM MEDICAL LTD | QWS12E | 05060434422309 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |