Narratives

Additional Manufacturer Narrative · 0

CONTINUATION OF D10: PRODUCT ID 9735821R, LOT NUMBER: UNKNOWN H3/H6: THE SYSTEM WAS SERVICED IN THE FIELD AND THE CAMERA WAS REPLACED. CODES B01, C08, AND D02 APPLY. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3: ANALYSIS WAS PERFORMED FOR PRODUCT: 9735821R, LOT NUMBER: P901405. IT WAS REPORTED THAT THE RETURNED POSITIONING SENSOR UNIT (PSU) CONTAINED SCRATCHES ON THE HOUSING LENSES. A CHECK OF THE EVENT LOG SHOWED INTERMITTENT FIRMWARE INCOMPATIBILITY. THERE WAS ALSO A BATTERY VOLTAGE LOW MESSAGE ALONG WITH BUMP DETECTED AND STORAGE TEMPERATURE EXCEEDED MESSAGES. THE PSU FAILED AN ACCURACY TEST (AAK) AT .496MM WITH A PASSING THRESHOLD OF .250MM. ALREADY REPORTED CODES B01, C08 AND D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT UPON BOOTING UP THE SYSTEM, THE SYSTEM GAVE LOCALIZER FAULTED MESSAGE. IT WAS CONFIRMED VIA THE NETWORK DEVICE INTERFACE (NDI) TOOLBOX THAT THE BUMP AND TEMPERATURE STATUSES WERE TRIGGERED. IT WAS ALSO NOTED THAT THE CAMERA CLIP WAS ALSO DAMAGED. THERE WAS NO PATIENT INVOLVEMENT.