FDA Adverse Event Injury Summary report: N

UNKNOWN SALINE IMPLANTS

MDR report key: 21978802 · Received May 7, 2025

Report

Report Number
1645337-2025-04937
Event Type
Injury
Date Received
May 7, 2025
Date of Event
January 8, 2025
Manufacturer
MENTOR TEXAS
Product Code
FWM
PMA / PMN Number
UNK
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. SINCE NO LOT NUMBER WAS PROVIDED, NO MANUFACTURING RECORD EVALUATION REVIEW COULD BE PERFORMED. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: BREAST PROSTHESIS DEFLATION. D4: UDI: AS THE CATALOG/MODEL NUMBER WAS NOT PROVIDED, THE (01)GTIN IS NOT AVAILABLE. MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON 02-JUL-2025, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON 15-JUL-2025, MENTOR COMPLETED THE INVESTIGATION ON THE SUSPECT MEDICAL DEVICE. DEVICE EVALUATION SUMMARY: ACCORDING TO THE INFORMATION RECEIVED, THE PATIENT EXPERIENCED A DEFLATION IN THE BREAST IMPLANT. THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION, LEAK TESTING, AND MICROSCOPIC EXAMINATION OF THE RETURNED DEVICE. DURING VISUAL EVALUATION, NO APPARENT DAMAGE OR VISUAL ANOMALIES WERE OBSERVED ON THE RETURNED DEVICE. LEAK TESTING WAS PERFORMED, IN ACCORDANCE WITH MENTOR PROCEDURES, AND A TEAR WAS OBSERVED AT THE UNION BETWEEN THE PATCH AND SHELL. MICROSCOPIC EXAMINATION WAS PERFORMED, AND THE CAUSE OF THE TEAR COULD NOT BE IDENTIFIED. AS THE AUTHORIZATION FORM FOR EXAMINATION WAS NOT RECEIVED, THE SHELL THICKNESS COULD NOT BE MEASURED. THE EVALUATION WAS LIMITED TO NON-DESTRUCTIVE TESTING. ALTHOUGH NO CONCLUSION COULD BE REACHED ON THE CAUSE OF THE REPORTED EVENT, THE INSTRUCTIONS FOR USE CONTAIN THE FOLLOWING CAUTION: CAUSES OF DEFLATION OF SALINE IMPLANTS INCLUDE BUT ARE NOT LIMITED TO THE FOLLOWING EVENTS: CREASE FOLD FAILURE, INTRAOPERATIVE OR POSTOPERATIVE TRAUMA, EXCESSIVE STRESSES OR MANIPULATIONS AS MAY OCCUR DURING NORMAL, DAILY ROUTINES INCLUDING CUSTOMARY AND PURPOSEFUL TRAUMA SUCH AS THAT WHICH CAN OCCUR DURING VIGOROUS EXERCISE, ATHLETICS, ETC. AS PART OF MENTOR'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED ACCORDING TO APPROVED SPECIFICATIONS. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW. ADDITIONAL COMPLAINT INFORMATION MONITORING FOR POTENTIAL SAFETY SIGNALS IS CONDUCTED THROUGH COMPLAINT TRENDING AS PART OF POST-MARKET SURVEILLANCE. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT AN UNSPECIFIED BREAST SURGERY WITH AN UNSPECIFIED MENTOR SALINE BREAST PROSTHESIS THAT DEFLATED POST IMPLANTATION. DEFLATION OF THE PATIENT¿S RIGHT BREAST PROSTHESIS WAS REPORTED. AS A RESULT, THE PATIENT UNDERWENT BILATERAL EXPLANTATION AND REPLACEMENT WITH MENTOR MEMORYGEL BREAST IMPLANT 325CC GEL BREAST PROSTHESES ON (B)(6) 2025.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1871678 UNKNOWN SALINE IMPLANTS PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE FWM MENTOR TEXAS

Patients

Seq Age Sex Outcome Treatment
1 62 YR Female Required Intervention