FDA Adverse Event Injury Summary report: N

TUFF IMPLANT

MDR report key: 21975990 · Received May 7, 2025

Report

Report Number
3011390931-2025-00314
Event Type
Injury
Date Received
May 7, 2025
Date of Event
April 14, 2025
Report Date
May 7, 2025
Manufacturer
NORIS MEDICAL LTD
Product Code
DZE
UDI-DI
07290108812176
PMA / PMN Number
K140440
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
DENTIST
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THE DHR WAS REVIEWED FOR LOT#9001144 AND THERE WAS NO EVIDENCE DISCOVERED TO INDICATE THAT A PRODUCT DEFECT OR NON-CONFORMITY CONTRIBUTED TO THE ISSUE. THE INVESTIGATION FOR THIS EVENT HAS NOT IDENTIFIED ANY SIGNALS INDICATING POTENTIAL NON-CONFORMANCES AFFECTING THE MANUFACTURING AND STERILIZATION PROCESSES. OUR MANUFACTURING SYSTEM ASSURES, THAT PRODUCTION AND PROCESS CONTROLS ARE IN PLACE TO ENSURE THAT BATCHES CONFORM TO THE APPLICABLE SPECIFICATIONS BEFORE THEY ARE DISTRIBUTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED WHICH INDICATES THAT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT, AN ADDITIONAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 0

THE HEALTHCARE PROFESSIONAL REPORTS THE IMPLANT WAS INSERTED (B)(6) 2025 IN THE PATIENT'S MOUTH. ON (B)(6) 2025, LOSS OF OSSEOINTEGRATION WAS REPORTED. ACCORDING TO THE INFORMATION, THE PATIENT IS SMOKER. OBSERVED DURING THE EVENT: BONE LOSS. THE DEVICE WAS FORWARDED TO THE MANUFACTURER. NO PATIENT OPERATIVE OR POST-OPERATIVE COMPLICATIONS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1831883 TUFF IMPLANT TUFF DENTAL IMPLANT DZE NORIS MEDICAL LTD NM-F4210 9001144 07290108812176

Patients

Seq Age Sex Outcome Treatment