FDA Adverse Event Malfunction Summary report: N

EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE

MDR report key: 21975310 · Received May 7, 2025

Report

Report Number
9610595-2025-07927
Event Type
Malfunction
Date Received
May 7, 2025
Date of Event
April 30, 2025
Report Date
May 7, 2025
Manufacturer
AIZU OLYMPUS CO., LTD.
Product Code
FDS
UDI-DI
04953170305276
PMA / PMN Number
K131780
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
003

Narratives

Additional Manufacturer Narrative · 0

THE DEVICE WAS RETURNED TO OLYMPUS FOR INSPECTION. A ROOT CAUSE COULD NOT BE IDENTIFIED. BASED ON THE RESULTS OF THE INVESTIGATION, IT IS LIKELY THE FOLLOWING LED TO THE MALFUNCTION: AIR/WATER-CYLINDER (TUBE) HAS FOREIGN OBJECTS, AIR/WATER-TUBE HAS FOREIGN OBJECTS, J-TUBE HAS FOREIGN OBJECTS AND DUE TO CLOGGING OF J-TUBE IN CONTROL UNIT, WATER CANNOT BE SUPPLIED. THE USE OF PRODUCTS THAT CONTAIN SILICONE CAN CAUSE DEPOSITS OF WHITE FOREIGN MATERIAL. SILICONE IS FOR EXAMPLE INCLUDED IN SIMETHICONE, A DEFOAMING AGENT, LUBRICANTS AND SO ON. IF THE CUSTOMER USED THEM, THERE WAS RISK THAT FOREIGN MATERIAL WAS REMAINED IN THE CHANNEL EVEN IF THE REPROCESSING WAS CONDUCTED IN ACCORDANCE WITH IFU. SIMETHICONE AND PETROLEUM/OIL/SILICONE-BASED LUBRICANTS ARE NON-WATER SOLUBLE AND THUS NOT RECOMMENDED FOR USE BY OLYMPUS. THE MOST PROBABLE CAUSE OF THE IDENTIFIED FAILURES "J-TUBE HAS FOREIGN OBJECTS", "AIR/WATER-CYLINDER (TUBE) HAS FOREIGN OBJECTS" AND "AIR/WATER-TUBE HAS FOREIGN OBJECTS." ARE CAUSE TRACED TO FAILURE TO FOLLOW INSTRUCTIONS. THE EVENT CAN BE PREVENTED BY FOLLOWING THE INSTRUCTIONS FOR USE WHICH STATE: NOTHING OTHER THAN STERILE WATER SHOULD BE USED FOR AIR/WATER FEEDING. NO ADDITIVES SHOULD BE PUT INTO THE STERILE WATER. NON-STERILE WATER MAY CAUSE PATIENT CROSS-CONTAMINATION AND/OR INFECTION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. OLYMPUS WILL CONTINUE TO MONITOR FIELD PERFORMANCE FOR THIS DEVICE.

Description of Event or Problem · 0

IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE FLEX VIDEO SCOPE EXHIBITED AIR/WATER-CYLINDER (TUBE) HAS FOREIGN OBJECTS, AIR/WATER-TUBE HAS FOREIGN OBJECTS, J-TUBE HAS FOREIGN OBJECTS AND DUE TO CLOGGING OF J-TUBE IN CONTROL UNIT, WATER CANNOT BE SUPPLIED. THERE WAS NO PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1513543 EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE GASTROINTESTINAL VIDEOSCOPE FDS AIZU OLYMPUS CO., LTD. GIF-HQ190 04953170305276

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown