IC-BIPOLAR HEAD COCRMO Ø 28 / 48 MM
Report
- Report Number
- 3012523063-2025-00044
- Event Type
- Injury
- Date Received
- May 7, 2025
- Date of Event
- March 31, 2025
- Report Date
- May 7, 2025
- Manufacturer
- IMPLANTCAST GMBH
- Product Code
- KWY
- PMA / PMN Number
- K191569
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
IT WAS REPORTED TO THE IMPLANTCAST GMBH THAT AN IC-BIPOLAR HEAD COCRMO Ø 28 / 48 MM DISLOCATED FROM THE NATURAL ACETABULUM. DURING AN ATTEMPT TO REPOSITION IT IN A CLOSED REDUCTION, THE IC-BIPOLAR HEAD THAN DISCONNECTED FROM THE FEMORAL HEAD. OPTICAL EXAMINATION OF THE EXPLANT WAS NOT POSSIBLE, AS IT WAS NOT PROVIDED TO THE IMPLANTCAST GMBH. THE MANUFACTURING DOCUMENTS AND DESCRIPTION OF THE SURGICAL TECHNIQUE WERE CHECKED, NO DEVIATIONS WERE FOUND. IT IS POSSIBLE THAT THE DISLOCATION WAS CAUSED BY TOO LITTLE LIGAMENT TENSION, WHICH ALLOWED THE IC-BIPOLAR HEAD TO SLIP OUT OF THE JOINT CAPSULE. HOWEVER, NO INFORMATION IS AVAILABLE TO SUPPORT THIS HYPOTHESIS. IN CASE A CLOSED REDUCTION IS PERFORMED AFTER DISLOCATION OF THE IC-BIPOLAR HEAD COCRMO FROM THE NATURAL ACETABULUM, THERE IS A RISK OF SEPARATION OF THE IC-HEAD FROM THE IC-BIPOLAR HEAD COCRMO. ALL IN ALL NO DEFINITIVE CONCLUSION CAN BE DRAWN AS TO WHY THE IC-BIPOLAR HEAD DISLOCATED. HOWEVER, IT IS ASSUMED THAT THE DISCONNECTION IS DUE TO THE CLOSED REDUCTION THAT WAS PERFORMED AFTERWARDS. THE EVENT WAS ASSIGNED TO THE HAZARD I "DISLOCATION / INSTABILITY" AND THE HAZARD II "SEPARATION /DISCONNECTION OF THE FEMORAL HEAD OUT OF BIPOLAR HEAD" IN THE ASSOCIATED RISK MANAGEMENT OVERVIEW.
THE FOLLOWING EVENT WAS REPORTED TO IMPLANTCAST GMBH: "DURING THE CLOSED REDUCTION OF THE BIPOLAR HEAD UNDER ANAESTHESIA, THE STEEL HEAD DISCONNECTED FROM THE BIPOLAR HEAD WITHOUT THE SNAP RING COMING LOOSE." REMARK IC-GMBH: THE TERMS DISCONNECTION AND DISLOCATION ARE USED AS FOLLOWS DISLOCATION: DUE TO THE DESIGN OF THE BIPOLAR HEAD IT IS POSSIBLE THAT THE WHOLE SYSTEM (THE BIPOLAR HEAD AND THE FEMORAL HEAD) CAN DISLOCATE OUT OF THE NATURAL ACETABULUM AS A WHOLE. DISCONNECTION: IT IS ALSO POSSIBLE THAT THE INDIVIDUAL COMPONENTS OF THIS SYSTEM DISCONNECT, RESULTING IN A SEPARATION OF THE BIPOLAR HEAD AND THE FEMORAL HEAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1872812 | IC-BIPOLAR HEAD COCRMO Ø 28 / 48 MM | FEMORAL HEAD | KWY | IMPLANTCAST GMBH | 21510048 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Female | Required Intervention |